Tools


Site Identifier

Brasil

BEP 2000 Advance System

Where convenience begins

Instrument 
Number of plates:Up to 4 plates at once
Number of samples Max100 tubes, up to max. φ 16 mm, length up to 200 mm
Number of reagentsUp to 4 reagent bottles in different sizes to be placed in special racks for Siemens reagents
Pre-dilution positionsPositions for up to 4 pre-dilution plates
Number of tipsUp to 5 tip racks for 300 μl or 1100 μl disposable tips
LoadingContinuous loading of plates, samples, reagents and tips
Photometer 
Spectral range400 - 700 nm
Dynamic range-0.100 to 3.000 O.D.
Accuracy+/- 0.005 or 2.5%
Linearity0 - 2.000 O.D. +/- 1%
DetectionPhoto diode
Reading time< 15 seconds
Read modesOD and Kinetic mode
FiltersUp to 8 positions
Pipetting System 
PipettorLiquid pipettor for disposable tips
Liquid level detectionStandard setting
Volumes min. / max.Small tips: 10–300 μL / large tips: 1000 μL
Accuracy<-15% at 25 μL (aqua)
<-5% at 100 μL (aqua)
Precision<5% CV at 25 μL
<2.5% CV at 100 μL
FeaturesTip detection, mixing, multi-dispensing mode
Incubation 
Capacity4 independent chambers
Temperature rangeMinimum temperature = RT + 5 °C
Maximum temperature = 50 °C
Accuracy+/- 1 °C mean of plate
Uniformity+/- 0.7 °C across plate
Washing 
CapacityMax. 3 positions wash buffer + 1 distilled water
Wash head1 x 8 with verify option
Dispense volume200 - 2500 μl/well
Precision+/- 5% CV at 300 μl
Residual volume< 2.5 μl in U-shaped bottom
< 4 μl in flat bottom
Fluid alarmsLow reagent, waste full
Features speedsSweeps, soak, purge, top & bottom wash, variable pump
Bar Code 
 Bar code identification of samples, reagents and test plates possible
Personal computer 
 Microsoft Windows® - compatible
Software features 
Operating systemMicrosoft Windows® 95, Microsoft Windows® 2000 or Microsoft Windows® 7
InterfacesASTM interface
Flexible ASCII import of worklist sample ID's
Time Scheduling Schedules4 plates with user options
Multiple assays per plateUp to 12 assays per plate
Export optionsFlexible ASCII export and report
QualitativeUser definable result classes
QuantitativeLinear to quadratic regression, sigmoidal, and many more
Language supportMultiple languages possible
QA analysisMean, SD, CV, Standard Error and Levey-Jennings
Power Supply 
Universal a.c. input100 - 260 VAC, 47 - 63 Hz
Typically max. 500 VA
Main fuses4 A T
Dimensions 
W x D X H114 cm x 156 cm x 100 cm
(Depth with waste bag and completely opened drawer)
Weight
 
 130 kg
Noise 
 67 dB (distance 1 m)
Environmental conditions 
 This standard applies equipment designed to be safe at least under the following conditions:
  indoor use
altitude up to 2000 m
temperature 5 °C to 40 °C
maximum relative humidity 80% for temperatures up to 31 °C decreasing linearly to 50% relative humidity at 40 °C
mains supply voltage fluctuations up to +/- 10% of the nominal voltage
applicable rated pollution degree
 Operating conditions 
 15 °C to 25 °C

 

O produto/recurso exibido nesta página Web não está comercialmente disponível em todos os países. Por razões de regulamentação, sua disponibilidade futura não pode ser garantida. Por favor, entre em contato com a organização da Siemens local para maiores detalhes.

Esta informação foi útil?
Agradecemos sua resposta
Detectamos números e/ou o símbolo "@" em seu comentário. Se estiver tentando inserir um número de telefone ou um endereço de email para que possamos contatá-lo, por favor entre em contato conosco por meio do nosso formulário de email.

Nota: para obter uma resposta direta e pessoal, por favor entre em contato conosco por meio do nosso formulario de email.

Obrigado